Equipment Remanufactured Systems Remanufactured OEC 9400 C-Arm
Remanufactured OEC 9400 C-Arm E-mail

ge-9400

All OEC-9400’s go through a multistep remanufacturing process.

Pain Management, Orthopedics, Rehabilitation Services are just a few of the applications utilizing the remanufactured OEC-9400. For barium swallows a VCR or Digital Video Recorder can be utilized for better record keeping and post study analysis.



THE PROCESS
  • All defective parts and covers are repaired or replaced
  • Vertical lift assembly is tested for drift and adjusted/replaced if necessary
  • All bearings are inspected, cleaned and re-lubricated, or replaced as neccessary
  • New cable pushers installed
  • All locks and brake assemblies are inspected, cleaned, and repaired or replaced as needed
  • New main connecting power and cable installed to include new low voltage and High Voltage cables
  • All boards are removed from main chassis and inspected for cold solder joints
  • Steering is tested and adjusted
  • Foot switch and hand switch provided on each unit
  • A complete set of operator and service manuals are provided
  • All covers are sent to the prep and paint area for cosmetic refinishing

Calibration and Final Testing

  • Image intensifier and X-ray tube are tested for stability and balance
  • Image intensifier is tested for resolution and gain to be within OEM specifications
  • X-ray tube bearings are tes noise and coast time
  • X-ray tube filaments and stator windings are tested
  • X-ray tube radiation output is verified to be within OEM specifications
  • X-ray generator high voltage tested
  • Every kVp and mA station tested fro accuracy
  • Image system tested and set for ABS sensing area
  • Maximum dose rate is set in compliance with government regulations
  • New or used Vidicon camera tube calibrated to OEM specifications
  • C-arm is tested for current leakage
  • Vertical lift assembly is tested for drift and adjusted/replaced as necessary with new style mechanism
  • All monitors are aligned, resized centered and focused
  • All power supplies are tested and calibrated or replaced as necessary
TECHNICAL SPECIFICATIONS
X-Ray Generator
  • High frequency 2.5 kHz
  • 7.5 KW Full-Wav
  • Up to 120kVp, up to 100 mA for radiographic exposures
  • Fluro boost/ pulsed fluro capability
    X-Ray Source
  • Rotating anode
  • Focal spots: 0.3 mm and 1.0 mm
  • Anode heat capacity: 300,000 HU
    Fluoroscopy Mode
  • Focal spot : 0.3 mm
  • kVp range: 40 - 120 kVp
  • mA range 0.2 - 5.0
  • PHYSYICAL SPECIFICATIONS Depth of arc: 23”, movement: 111º
  • Travel: - Verticla: 18” motorized. - Horizontal: 8”
  • L-arm rotation: ± 360º motorized
  • Reversible C-arm: Manual
    Input Power
  • 100/120 VAC, 15 AMP
  • 200/220/240 VAC, 12 AMP
  • 50 Hz (optional)
    Main C-arm Frame (stored)/ Monitor Cabinet Dimensions
  • Length: 74.5” (189 cm)/ 32”
  • Height 66.2” (168 cm)/ 65”
  • Depth: 34” (86 cm)/ 27” (63 cm)
  •